トップページ > 医療機器SAR測定 > ELIT 3.0



ELIT 3.0


The Elliptical Implant Test (ELIT) phantoms have been developed to generate uniform exposure for implants in the MITS systems and enable automated testing with DASY52NEO . They allow the exposure of implants with long leads at constant amplitude and phase (MRI).

The phantoms meet the specifications of the current ISO/IEC AIMD JWG Test Plan.


Material

Transparent Plexiglas (PMMA)

Liquid Compatibility

Compatible with sugar and oil based tissue simulating media.

Not compatible with DGBE based liquids

Dimensions

Length

500mm

Height

300mm

Width

90mm

Submerged depth
of implant

84mm

Wall Thickness

11mm

Racetrack Material

FR-4 Glass reinforced epoxy

Racetrack
Dimensions

Length

336mm

Height

131.7mm

Width x 1 Track

2.1mm

Width x 3 Track

26.3mm

Track spacing

10mm

Amplitude and phase
along top of racetrack

Etan 75 ± 5% of Emax

Phase 0 ± 8°

 Complete ELIT3.0 Phantom
 with racetrack

 Removable racetrack

 Close-up of the racetrack


詳細やお見積もり、お問合せはこちらまで